Radiofrequency catheter ablation for the treatment of cardiac arrhythmias [
In a previous work [
This nonrandomized study involved 1028 consecutive patients (pts) (1145 procedures) who underwent CA procedures from January 2007 to December 2016. Overall, 350,000 residents refer to the EP Lab of our hospital and the total number of CA procedures per year is approximately 140. We report the data on all CA procedures for each year; in 2007 and 2008, some of the procedures were performed by using only fluoroscopy, whereas in subsequent years (from 2009 to 2016) all the procedures were performed by also using a nonfluoroscopic mapping system (NMS). We would like to emphasize that the data for each year were compared with those for each other year and, in particular, with the data from 2007, when the CA procedures were mainly performed with fluoroscopy. We carried out a detailed subanalysis only for AF CA and SVT CA.
In 2001, we started using a single NMS (CARTO, Biosense-Webster, Diamond Bar, CA, USA, Unix Version) in CA procedures for the treatment of complex arrhythmias (atrial fibrillation, atypical atrial flutter, atrial tachycardia, and ventricular tachycardia). In 2007, the new CARTO XP Version allowed us to integrate CT/MRI images of the cardiac chambers with the electroanatomic maps. As a result, images of the left atrium (LA) were integrated during AF ablation or atypical atrial flutter ablation procedures. The following year (2008), we started to use a second NMS (EnSite NavX™ St. Jude Medical, St Paul, MN, USA) for SVT CA (AVNRT, AVRT, atrial tachycardia, atrial flutter). Since October 2008, in agreement with our administration, all CA procedures, with the exception of His-bundle ablation procedures, have been performed by using one or the other of these two NMSs. It should also be specified that, in our EP Lab, all AF CA procedures have always been performed by using the CARTO system, and almost all SVT CA procedures have been performed by using the EnSite system. This use of a specific mapping system for a specific type of ablation procedure was for no other reason than our prior experience and confidence with the specific NMS. New updated hardware and software versions of these NMSs have been used since 2010 (CARTO3 and EnSite Velocity) and since 2015 (EnSite Precision). Fluoroscopy was performed by means of a Toshiba Radiographic/Fluoroscopic Unit (Infinix-I Series CAS, Toshiba Medical System Corporation, Japan). We utilized the same number of catheters during the procedures, regardless of whether fluoroscopy or an NMS was used. Four diagnostic catheters (3 quadripolar and 1 decapolar) and the ablation catheter were used for SVT CA procedures (right jugular vein and right femoral vein as access) and three diagnostic catheters (1 duo-decapolar, 1 quadripolar, and 1 circular mapping catheter) and the ablation catheter were used for AF CA procedures (right jugular vein and right femoral vein as access). From 2007 to 2012, all CA procedures were performed by an experienced operator and by a second operator who was finishing his period of training during the first 3 years of our study. Another two operators started their training in 2013 and 2016, respectively; they were first involved in SVT CA procedures and subsequently in more complex procedures, such as AF CA. During this last period an experienced operator left the lab. We consider an operator who has achieved the minimum procedural volume proposed by ACC/AHA/HRS 2015 Training Statement experienced [
In this study, we compared fluoroscopy time, total X-ray exposure, and the duration of each CA procedure. Fluoroscopy time (FT) was defined as the cumulative duration of fluoroscopy during the entire procedure, whereas the patient radiation dose was assessed as the recorded dose-area product (DAP). Procedure time (PT) was measured as the interval from the initial recording of intracardiac signals to the final ECG recording before the end of the procedure.
The definition of procedural success was based on a specific type of tachycardia, as is commonly the international criterion [
A Shapiro-Wilk normality test was performed on continuous variables to examine the normal distribution of each variable. To compare all continuous variables and to test median values between groups, nonparametric pairwise Wilcoxon Rank Sum Tests were performed, with Benjamini-Hochberg correction for multiple testing. For categorical variables, analysis was performed between groups by means of the
Box and Whisker Plots were used to depict the trends in the parameters DAP and FT year by year for all CA procedures.
To assess the impact of experience of the systematic use of NMSs year by year, we reported the median values of the parameters DAP, FT, and PT as histograms, with the main statistically significant differences (
Regarding follow-up and long-term effects, Kaplan-Meier analysis of Disease-free Survival and the Log-Rank Test for comparison between groups were used. Statistical analyses were conducted by means of R software, Version 3.4.3 [
Our study included 1028 pts (620 males, 408 females) (1145 procedures) who underwent CA from January 2007 to December 2016. Table
Clinical characteristics of the population of this study.
TOTAL | 2007 | 2008 | 2009 | 2010 | 2011 | 2012 | 2013 | 2014 | 2015 | 2016 | |||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
# | % | # | # | # | # | # | # | # | # | # | # | ||
PROCEDURES | 1145 | 100 | 110 | 119 | 102 | 117 | 122 | 125 | 108 | 104 | 114 | 124 | |
| |||||||||||||
AGE | Mean [years] | 55.7 | - | 53.9 | 53.3 | 57.5 | 54.7 | 56.5 | 54.4 | 55.3 | 57.1 | 56.5 | 58.1 |
Standard deviation [years] | 16.1 | - | 16.2 | 16.8 | 14.4 | 16.7 | 16.22 | 16.1 | 16.6 | 17.3 | 15.6 | 14.3 | |
| |||||||||||||
SEX | F | 449 | 60.8 | 39 | 55 | 44 | 44 | 45 | 47 | 43 | 39 | 57 | 36 |
M | 696 | 39.2 | 71 | 64 | 58 | 73 | 77 | 78 | 65 | 65 | 57 | 88 | |
| |||||||||||||
HEART DISEASE | None | 551 | 48.1 | 61 | 61 | 47 | 63 | 58 | 61 | 53 | 42 | 59 | 46 |
Hypertensive | 335 | 29.3 | 23 | 27 | 28 | 25 | 31 | 44 | 29 | 37 | 36 | 55 | |
Valvular | 74 | 6.5 | 8 | 15 | 12 | 5 | 10 | 5 | 2 | 6 | 3 | 8 | |
Ischemic | 85 | 7.4 | 10 | 6 | 8 | 11 | 7 | 4 | 9 | 14 | 11 | 5 | |
Dilated cardiomyopathy | 54 | 4.7 | 6 | 5 | 4 | 8 | 12 | 3 | 8 | 1 | 1 | 6 | |
Congenital | 14 | 1.2 | 0 | 5 | 2 | 1 | 0 | 1 | 2 | 1 | 1 | 1 | |
Other | 32 | 2.8 | 2 | 0 | 1 | 4 | 4 | 7 | 5 | 3 | 3 | 3 | |
| |||||||||||||
LVEF | < 35% | 46 | 4.0 | 6 | 1 | 5 | 5 | 5 | 2 | 8 | 5 | 3 | 6 |
> 55% | 917 | 80.1 | 89 | 85 | 78 | 97 | 97 | 106 | 77 | 88 | 97 | 103 | |
35-45% | 42 | 3.7 | 5 | 5 | 4 | 7 | 8 | 2 | 4 | 2 | 3 | 2 | |
45-55% | 140 | 12.2 | 10 | 28 | 15 | 8 | 12 | 15 | 19 | 9 | 11 | 13 | |
| |||||||||||||
ARRHYTHMIA | AF | 378 | 33.0 | 35 | 29 | 27 | 39 | 48 | 50 | 34 | 36 | 34 | 46 |
SVT | 435 | 38.0 | 49 | 54 | 46 | 46 | 33 | 48 | 39 | 31 | 45 | 44 | |
AFL/AT | 244 | 21.3 | 25 | 32 | 24 | 18 | 27 | 23 | 22 | 27 | 21 | 25 | |
VT | 88 | 7.7 | 1 | 4 | 5 | 14 | 14 | 4 | 13 | 10 | 14 | 9 |
F, female; M, male; LVEF, left ventricular ejection fraction; AF, atrial fibrillation; SVT, supraventricular tachycardia; AFL, atrial flutter; AT, atrial tachycardia; VT, ventricular tachycardia
Figure
The Box and Whisker Plots show the trend of fluoroscopy time (FT), expressed in minutes [min], in all CA procedures.
The Box and Whisker Plots show the trend of dose-area product (DAP), expressed in [Gy
During this period, the maximum reduction in exposure and PT was achieved in the year 2012. The median fluoroscopy time fell by 84% to a value of 4.7 min (95% CI [3.7, 6.0]) (P<0.001), with a concomitant reduction in total X-ray exposure of 85% to a value of 8.48 Gy
This can be explained by the systematic use of NMSs and the increased experience of the operators. After 2012, the entry of new operators with no experience in NMSs led to a slight increase in exposure to fluoroscopy, confirmed by the longer PT.
The percentage of CA procedures performed only with fluoroscopy was 44% in 2007 and 34% in 2008; in the following years, no CA procedures were performed only with fluoroscopy.
During the period of our study, 373 AF CA procedures (320 pts) were performed, all of which involved using a single NMS (CARTO). From 2007 to 2016, fluoroscopy time was reduced by 74%, from 49.1 min (95% CI [41.4, 56.6]) to 13 min (95% CI [10.8, 14.6]) (P<0.001) and total X-ray exposure was reduced by 69%, from 138 Gy
During this period, the maximum reduction in exposure and PT was achieved during the year 2012. The median fluoroscopy time fell by 85% to a value of 7.5 min (95% CI [6.0, 9.7]) (P<0.001) with a concomitant reduction in total X-ray exposure of 86%, to a value of 18.9 Gy
Figure
Histograms show the trend of median values of fluoroscopy time (FT), expressed in minutes [min], dose-area product (DAP), expressed in [Gy
A total of 388 SVT CA procedures (379 pts) were performed during the study period (2007-2016). Like the AF CA procedures, the SVT CA procedures were carried out mainly using a single NMS (EnSite system). During the first period of the study (2007-2011) fluoroscopy time decreased by 93%, from 16.6 min (95% CI [12.2, 21.8]) to 1.2 min (95% CI [0.1, 2.5]) (P<0.001); this was mirrored by a corresponding 92% reduction in DAP, from 12.70 Gy
After 2011, the entry of new operators with less experience of NMSs led to a slight increase in exposure to fluoroscopy, the 2016 FT value being 3.0 min (95% CI [1.7, 5.7]) (P<0.001) and the DAP value being 5.94 Gy
Figure
Histograms show the trend of median values of fluoroscopy time (FT), expressed in minutes [min], dose-area product (DAP), expressed in [Gy
The postprocedural echocardiogram was normal in almost all patients after CA procedures. However, we had 15 cases of pericarditis and 8 cases of pericardial effusion after ablation (one after transseptal puncture; the procedure was stopped). Furthermore, we had 13 cases of postprocedural femoral hematoma and 2 of venous-arterial fistula. It must be stated, however, that the nature of these complications was not different from the types of complications encountered by our group in the years of work in our EP Lab before the study.
Regarding the procedural follow-up and long-term effects, Kaplan-Meier analysis of disease-free survival was used for both AF CA and SVT CA procedures.
Figure
Kaplan-Meier estimates of disease-free survival with 95% confidence interval for all AF CA and SVT CA procedures, with also the table of number of patients at risk reported as number and percentage.
Kaplan-Meier estimates of disease-free survival with 95% confidence interval for AF CA procedures for each year, with also the table of number of patients at risk reported as number and percentage for each year.
Kaplan-Meier estimates of disease-free survival with 95% confidence interval for SVT CA procedures for each year, with also the table of number of patients at risk reported as number and percentage for each year.
The use of mapping systems was originally limited to the treatment of complex arrhythmias, but it was subsequently shown that their use also for simple arrhythmias may result in simpler and quicker procedures [
An advantage of the use of a mapping system is the significant reduction in fluoroscopy exposure for both patients and operators [
Our study demonstrated that the extended use of an NMS dramatically reduced radiation exposure in an unselected population of patients, with only slight fluctuations when two trainee operators joined our EP Lab team in 2013 and in 2016.
The initial decrease in radiation exposure was mainly due to the fact that, from January 2007 to October 2008, just under half of the CA procedures were performed by using fluoroscopy as the only imaging modality; subsequently, the systematic use of NMSs was adopted in our EP Lab. Moreover, improvements in NMS software allowed the operator to use less fluoroscopy to check catheter movements, as seen for AF CA procedures. Indeed, after the image integration of the LA, the possibility of visualizing the circular mapping catheter by means of CARTO 3 permitted a further reduction in fluoroscopy in comparison with the use of the CARTO XP Version. Similarly, the implementation of the new software version of EnSite Velocity led to a reduction in fluoroscopy use. In addition, the decrease in radiation exposure can be ascribed to the progressive change of mindset of the electrophysiologist, who is increasingly aware that the use of fluoroscopy during the procedure is sometimes superfluous. In fact, the only use of NMSs does not automatically reduce the radiation dose [
The data collected in these 10 years of activity at our EP Lab reveal that the systematic use of these new mapping systems is linked to operator learning and experience in a context where these systems are already used. Until 2012, all procedures performed with the aid of NMSs were performed by two operators, who had used these mapping systems from the beginning. Indeed, the maximum optimization of these techniques was reached in 2011 for SVT CA and in 2012 for AF CA. In 2013, an expert operator was replaced by a new operator, who began with the extensive use of NMSs in traditional procedures. At mid-year, the new operator began to take over from more experienced colleagues, even in the most complex procedures, AF CA.
If we consider the lower use of fluoroscopy, as well as a shorter procedure time, to be an indicator of operator experience of NMSs, we can see that the increase in the parameters of DAP, FT, and PT in SVT CA procedures in 2013, in comparison with 2011, reflects the entry of an unskilled operator. It can also be inferred, however, that, within one year, this operator gained experience with traditional procedures, as indicated by the reduced use of fluoroscopy. In 2016, a second new operator, with less experience, started to perform SVT CA procedures, which is again reflected by an increase in the use of fluoroscopy and in procedure time.
After considering the possible factors involved in this significant radiation reduction, an important question remains: why has our EP Lab, despite this 10-year systematic use of NMSs, not succeeded in achieving zero fluoroscopy use? In our opinion, part of the reason for this is that NMNs were created and developed in order to provide a 3D image of the heart for the interpretation of complex arrhythmias. Indeed, the NMS is a dedicated and highly focused mapping tool which does not consider the context of the heart in terms of the organs and structures that surround it (lungs, great vessels, spine, bronchial tree, diaphragm, etc.). By contrast, fluoroscopy provides a less detailed two-dimensional, but bigger, picture. Furthermore, if we subdivide the CA procedure into four parts (venous access, catheter positioning and possible transseptal puncture, electrophysiological study, and ablation, in that order), we can appreciate that the mapping system can significantly reduce radiation exposure in the last two phases, but less in the first two, especially if the operator is less experienced. This is because the mapping system does not allow the operator to visualize wires, long sheaths, transseptal needle, catheter curves, etc. In practical terms, fluoroscopy has its uses throughout the procedure, because of its broader scope.
An extended use of NMSs is associated with increased procedure costs. Nonetheless, this increase is likely compensated by the decrease over the years of radiation induced malignancies. In our previous work on NMSs in children [
This study has at least three limitations: first, it was not a randomized multicenter study; second, we did not analyze the cost/benefit effect of this extended use of NMSs and, third, we did not analyze the effect of the reduction in fluoroscopy on the specific risks of exposing patients and medical professionals to radiation.
Moreover, in this study we did not perform an operator-based analysis, describing for each of those involved, independently, trends over the years. This would have allowed identifying a cut-off number of procedures after which performance levels to that of the reference experienced operator.
The systematic use of an NMS dramatically reduces fluoroscopy time and total X-ray exposure in comparison with the use of fluoroscopy as the main guiding tool in CA procedures. In our experience, this reduction was mainly due to the daily use of NMSs, which enabled EP team members to develop their competence and confidence. We believe that the marked reduction in X-ray exposure reported in our “real-life” study provides a strong argument for the routine use of NMSs in CA procedures and justifies the increased cost. We must also consider that, given the ethical imperative to protect the health of patients and operators, treating cardiac arrhythmias without using an NMS is no longer justifiable. On the other hand, further technological commitment from the industry is needed, in order to overcome the limits of current mapping systems and to achieve total zero fluoroscopy.
The data used to support the findings of this study are available from the corresponding author upon request.
Present work was performed as part of the employment of the authors at their hospital institutions.
The authors declare that they have no conflicts of interest.
We would like to thank Mr. Bernard Patrick for proofreading and correcting the English of the paper.