Knowledge of the fate and transport of nanoparticles in the subsurface environment is limited, as techniques to monitor and visualize the transport and distribution of nanoparticles in porous media and measure their in situ concentrations are lacking. To address these issues, we have developed a light transmission and fluorescence method to visualize and measure in situ concentrations of quantum dot (QD) nanoparticles in variably saturated environments. Calibration cells filled with sand as porous medium and various known water saturation levels and QD concentrations were prepared. By measuring the intensity of the light transmitted through porous media exposed to fluorescent light and by measuring the hue of the light emitted by the QDs under UV light exposure, we obtained simultaneously in situ measurements of water saturation and QD nanoparticle concentrations with high spatial and temporal resolutions. Water saturation was directly proportional to the light intensity. A linear relationship was observed between hue-intensity ratio values and QD concentrations for constant water saturation levels. The advantages and limitations of the light transmission and fluorescence method as well as its implications for visualizing and measuring in situ concentrations of QDs nanoparticles in the subsurface environment are discussed.
Engineered nanomaterials such as carbon nanotubes, fullerenes, and metal- and semiconductor-based nanomaterials (e.g., quantum dots (QDs)) are revolutionizing science and engineering. Nanomaterials release into soil, air, and water resources following their life cycle or spills and their impacts on ecological systems are inevitable [
To investigate the fate and transport of nanoparticles in the subsurface environment, the common practice is to use laboratory-scale saturated or unsaturated column experiments, in which the nanoparticles are introduced from one end of the column under various solution chemistry and hydrodynamic conditions and are collected from the other end over time. Changes in the nanoparticle concentrations are usually obtained only from the end of the column with respect to time or pore volume eluted; these data are called breakthrough curves (BTCs) [
Among the various types of nanomaterials that can be selected for the study of transport in the subsurface environment, semiconductor QDs are important subjects of investigation for several reasons. QDs were developed in the 1980s, a decade before the development of other nanomaterials such as carbon nanotubes. QDs are essential to nanosciences, engineering, and technology and are used in biology, medicine, and industry due to their optical and electronics properties [
The objective of our research is to develop a nonintrusive visualization method capable of not only visualizing QDs but also providing in situ measurement of QDs’ concentration in variably saturated porous media. This paper presents the experimental apparatus and procedures used in a visualization method that combined both light transmission and fluorescence to visualize QDs in variably saturated porous media and discusses its potential application.
CdSe/ZnS QDs were produced according to the procedures previously developed by researchers [
Ottawa sand, a nonporous quartz silica sand of 0.7 mm (20/30 US std sieve size) grain size, was selected as soil porous medium because of its translucence and other optical properties and previous application in light transmission methods [
Color can be represented in two common formats: red, green, and blue (RGB) or hue, saturation, and intensity (HSI). In RGB color format, the amplitudes of red, green, and blue for a particular color can each range from 0 to 100 percent of full brilliance. In HSI format, hue is the color of an object, saturation is the amount of pure white in the color, and intensity represents the brightness of an object [
The LTFM experimental setup includes two-dimensional calibration cells packed with porous media, a fluorescent light source, a UV light, a lens, a camera, and image acquisition and analysis software (Figure
Experimental setup.
The fluorescent white light source was an Osram Planon (Osram, Munich, Germany) flat fluorescent light made of light-emitting diodes (LEDs) and producing a uniform high luminance from center to edge. The luminous flux was 20 lm and the luminous intensity was 0.3 cd. The diagonal length and the thickness of the Osram Planon were 26.4 cm and 1 cm, respectively. The fluorescent light source was mounted in the back of the experimental chamber. The UV light consisted of a fluorescent black light (Altman Stage Lighting Co., Inc. NY, USA) made of “Wood’s Glass,” that is, glass doped with nickel oxide, resulting in a purple color blocking all visible light above 400 nm. The UV light dimensions were 11 cm high × 14 cm wide × 125 cm long, with a power of 40 watts. The UV light was placed about 10 cm in front of the experimental chamber.
The camera used to record images of the transmitted and emitted lights under the different light exposures was a MicroPublisher 3.3 RTV (QImaging, Surrey, BC, Canada), which is a digital charge-coupled device (CCD) color camera with high-speed real time viewing (RTV). The MicroPublisher 3.3 RTV is a high resolution camera equipped with a 3.3 million pixel sensor system (2048 × 1536) that allows a region of interest (ROI) from 1 × 1 pixels up to full resolution. The MicroPublisher 3.3 RTV is able to display a full field of view with frame rates up to 30 fps. The camera was mounted on a tripod and located 0.5 m in front of the experimental chamber. The camera was connected through a digital interface IEEE 1394 FireWire to a computer equipped with MS Windows (Microsoft Corporation, Redmond, WA, USA) as the operating system. The image acquisition, processing, color format conversion, measurement, and analysis were performed using IPLab software (version Va.0.4, 2006) for Windows from Scanalytics, Inc. VA, USA.
Under the fluorescent light exposure, the transmitted light resulting from the variably saturated porous media containing QDs was recorded with the CCD camera in RGB format. The image was then converted from RGB to HSI format using the IPLab software; this created an image called the “intensity image” from which the intensity of light transmitted could be measured. This image is analyzed in order to visualize water in porous media and to quantify water saturations. Under the UV light exposure, the emitted light produced by the excited QDs present in the variably saturated porous media was recorded with the CCD camera in RGB format. The image was then converted from RGB to HSI format using the IPLab software to create an image, called the “hue image,” from which the hue (color) of the light emitted could be measured. This image was analyzed in order to visualize QDs in variably saturated porous media and to measure in situ QD concentrations.
To establish a calibration between hue values and QD concentrations and between intensity values and water saturations, a two-dimensional calibration system consisting of cells packed with sand porous media and known quantities of QD concentrations and water saturations was constructed. A wide range of QD suspension concentrations was derived from the green-emitting QD stock suspension. Six different concentrations of QD suspension—40, 100, 150, 200, 300, and 500 ppm—were prepared by dilution of the stock suspension of QDs using deionized water. Five different water saturations—0, 25, 50, 75, and 100%—were selected. Calibration cells were plastic disposable fluorometer cuvettes (Perfector Scientific, Atascadero, CA, USA). These cuvettes were made of methacrylate (useable wavelength range: 285 to 800 nm). The use of cuvettes made of methacrylate, therefore, ensures minimal wall effect of the cuvette on the visualization of QDs. Sand was packed into the cuvettes of internal dimensions 1.1 cm high × 1.1 cm wide × 4.5 cm long to create a two-dimensional porous matrix. The calibration cells were filled with sand to achieve porosity of 0.38 with all possible combinations of QD concentration suspensions and degrees of water saturation.
The calibration cells were then stacked and positioned horizontally along their length in the LTFM experimental system. The experimental procedure consisted of switching on the fluorescent light source to record the transmitted light through porous media while the UV light is switched off and then switching on the UV light to excite the QDs and record the emitted light of the QDs while the fluorescent light is switched off. The calibration cells were exposed successively in an alternating fashion to fluorescent light and UV light exposures; the resulting images from the transmitted and emitted light were analyzed by the imaging IPLab Software. For each calibration cell, the intensity value resulting from the “intensity image” obtained with the transmitted light and the hue value resulting from the “hue image” obtained with the emitted light were measured as the mean value in a rectangle containing 100 pixels. These average values were then used to establish calibration curves corresponding to water saturations and QD concentrations.
The sizes and surface charges of QDs in suspensions of different ionic strengths and pH were characterized through Dynamic Light Scattering (DLS) measurements of hydrodynamic diameter and zeta potential measurements using a NanoBrook 90Plus Zeta Particle Size Analyzer Brookhaven Instruments Corporation (Holtsville, NY). The stability of different concentrations of QDs under different ionic strength and pH conditions was assessed through monitoring the absorbance at 260 nm. The stability of the QDs was calculated as the ratio of the absorbance at 260 nm of the QD suspensions under various chemical conditions (i.e., ionic strength and pH) and the absorbance of the QD suspensions at ionic strength 5 × 10−4 M NaCl or pH 6.5 following the procedures of stability analysis of Lin et al., 2010. To assess the stability of QDs, the suspension ionic strength was adjusted with NaCl while the pH was adjusted using 0.1 M NaOH and 0.1 M HNO3. Four ionic strengths (5 × 10−4 M, 5 × 10−3 M, 5 × 10−2 M, and 0.5 M) and five pH values (3, 4.5, 5.5, 6.5, and 8) as well as pH 7 and deionized water were used.
The stability of QDs was impacted by the suspension ionic strength and pH. As the suspension ionic strength increased from 5 × 10−3 M NaCl to 0.5 M NaCl, the stability of QDs in suspension remained at or above 80%, an indication of limited aggregation of the QDs. As the suspension pH remained above 6, the stability of the QDs was above 80%, while as the pH suspension decreased from 6 to 3, the stability of the QDs decreased to 17%, leading to significant aggregation of the QDs. Ionic strength and pH impacted the zeta potential of QDs. The impact of ionic strength and pH on the stability of the QDs was limited as the QDs were originally coated by neutralized amphiphilic polymer (octylamine-modified polyacrylic acid). The addition of the polymer to the coating of QDs contributed to the steric repulsion forces. The presence of the polymer at the surface of QDs stabilizes them against aggregation. The average hydrodynamic diameters of QD aggregates were
The LTFM allowed the visualization of differences between the intensity and the hue among the calibration cells containing QDs in variably saturated porous media. These were differences resulting from transmitted light under fluorescent light exposure and emitted light under UV light exposure, respectively (Tables
RGB image and intensity image of the calibration cells under fluorescent light exposure.
QDS concentration (ppm) | Water saturation (%) | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
0% | 25% | 50% | 75% | 100% | ||||||
RGB |
Intensity |
RGB |
Intensity |
RGB |
Intensity |
RGB |
Intensity |
RGB |
Intensity | |
40 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
100 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
150 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
200 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
300 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
500 ppm |
|
|
|
|
|
|
|
|
|
|
RGB image and hue image of the calibration cells under UV light exposure.
QDS concentration (ppm) | Water saturation (%) | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
0% | 25% | 50% | 75% | 100% | ||||||
RGB |
Hue |
RGB |
Hue |
RGB |
Hue |
RGB |
Hue |
RGB |
Hue | |
40 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
100 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
150 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
200 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
300 ppm |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
500 ppm |
|
|
|
|
|
|
|
|
|
|
The intensity and hue values were obtained from the conversion of the images from RGB to HSI format using the IPLab software. The intensity values ranged from black, the lowest intensity value (12), to white, the highest value (160). The 1976 chromaticity diagram of the Commission Internationale de I’Eclairage was used to display the hue values. The hue ranged from 27° to 75° in the case of the saturated porous media and from 75° to 180° in the case of the unsaturated porous media. In each cell, the hue and intensity values were almost uniform and there were few differences within the cells.
QD concentrations in variably saturated porous media were determined by deriving relationships from the intensity and hue values resulting from transmitted light and emitted light, respectively, obtained from images of the calibration cells. Two calibrations were performed to determine the water saturation and the QD concentration. The first calibration established the relationship between the porous media water saturation of the calibration cells containing a wide range of QD concentrations and the intensity component of the transmitted light obtained by placing a fluorescent light behind the calibration cells. The second calibration assessed the relationship between the QD concentrations and the hue component of the emitted light of the calibration cells resulting from UV light exposure.
The light intensity of the calibration cells containing a wide range of QD concentrations in variably saturated porous media increased as water saturation increased. The mean intensity values calculated for different water saturations containing various QD concentrations are displayed in Figure
Intensity of the light transmitted through variably saturated porous media under fluorescent light exposure as a function of water saturation.
Light intensity and water saturation are uniquely related and can be estimated by
The hue values representative of the QD concentrations of the calibration cells showed two trends in reference to the water saturation scenarios. In saturated porous media conditions, an increase in QD concentration resulted in an increase in the hue value. In the case of unsaturated porous media conditions, on the other hand, an increase in the QD concentrations led to a decrease in the hue value when water saturation was constant. For the same QD concentration, the highest hue value belonged to the cell with the lowest water saturation (25% saturation), and as the water saturation of the calibrations cells increased, the hue value decreased. The discrepancy between saturated and unsaturated porous media conditions indicated that hue was also impacted by water saturation and therefore intensity. Thus, in the case of variably saturated porous media containing fluorescent nanoparticles (QDs), the light that was emitted was a function not only of the QD fluorescence (and therefore QD concentration), but also of the water saturation or air saturation (i.e., light intensity). The fluorescence of QDs in variably saturated porous media was affected by air saturation; therefore, adjusted hue values were calculated by taking the ratio of the hue value resulting from emitted light under UV exposure and the intensity value resulting from the transmitted light under fluorescent light exposure. The alternately switching on and off of the fluorescent and the UV lights allowed two independent hue and intensity images of the calibration cells to be obtained.
The hue value was not exclusively a function of QD concentrations; it depended on both QD concentrations and water saturations. To account for the effects of water or air saturation in porous media on the emitted fluorescence of QDs, and therefore on the hue detected in the calibrations cells with different QD concentrations, an adjusted hue value (
The adjusted hue values (
Quantum dots concentration versus hue-intensity ratio for constant water saturations in variably saturated porous media.
For constant water saturations, the QD concentration QD (for
When developing the relationships between
In summary, water saturation is determined first with (
A nonintrusive, highly spatial and temporal light transmission and fluorescence method (LTFM) was developed to visualize and measure in situ concentrations of QD nanoparticles in variably saturated porous media. The LTFM and its calibration procedure provide direct visualization and in situ measurements of QD concentration in variably saturated porous media compared to other methods that may use relative measurements. Although the application of LTFM is limited by the requirement to use translucent porous media, fluorescent nanoparticles, and two-dimensional porous media chambers, visualization of QDs and in situ measurements of QDs in variably saturated porous media by the LTFM allow a quantitative investigation of the mobility and transport of QDs in porous media in situ and in real time to be made, in contrast to traditional column experiments and derived breakthrough curves.
Unsaturated porous media conditions were found to affect the visualization of QDs since their fluorescence is impacted by the water saturation or air saturation; therefore, the hue values resulting from the fluorescence of QDs were adjusted with respect to the intensity values which represent the water saturation or air saturation. Water saturation was directly proportional to the intensity value. A linear relationship was observed between adjusted hue values (
The minimum of the QD nanoparticle concentration that could be detected by the LTFM was estimated at 5 ppm through extensive experimentation. The sensitivity and accuracy of the LTFM can be assessed through the range of
The Beer-Lambert law states that the intensity of the fluorescence is equal to the intensity of the adsorbed light multiplied by the fluorescence quantum yield [
As the QDs’ residency in the unsaturated porous media will also potentially influence the light transmission and scattering, it is important to consider where the QDs reside, that is, in the water, at the gas-water interfaces, or at the solid-water interfaces. Previous research has demonstrated that, in unsaturated porous media systems, the primary function of gas-water interfaces is the capture and retention of QDs, while the secondary function is postulated to be the potential transfer of previously captured QDs to solid-water interfaces ensuing their deposition at solid-water interfaces [
The LTFM may be applied to characterize the mobility and transport of QD nanoparticles in the saturated and unsaturated zones over scales ranging from
The LTFM may be used to detect light-induced reactive surface chemistry of QDs for environmental sensing implications that use photoluminescence and reactivity. As our research demonstrated the potential of the LTFM to detect and quantify the fluorescence emitted by QDs, changes in the chemical optical properties of reactive QDs as a result of chemical reactions can be detected and monitored in real time by the LTFM. Based on the physicochemical properties of reactive QDs, biochemical changes in the environment that reactive QDs encounter or are placed in may be sensed and monitored through chemical optical detection using the LTFM. Therefore, the coupling of the LTFM as a visualization method and reactive QDs as a smart nanoparticle-sensing technology can be applied to monitor physical, chemical, and biological changes that occur in environmental systems. It has potential applications in the fields of soil, water, and environment quality, as well as environmental remediation and protection of ecosystem and public health. However, the stability of reactive QDs in the environment and their interactions with surfaces and other biochemical compounds from soil, plants, and microorganisms should be further examined and controlled to ensure the integrity and functionality of reactive QDs as smart chemical sensors once they are released into the natural environment. In the natural environment, the LTFM and reactive QDs can be applied to study certain characteristics of natural soils, such as their biochemical characteristics, rather than the behavior of nanomaterials in natural soils due to the soil components that may mask any artificial nanoparticles present in soils. The coupling of the LTFM as a visualization method and reactive QDs as a smart nanoparticle-sensing technology may contribute to the potential applications of QD sensor technology for early detection and warning of the presence of contaminants in the environment and for monitoring of biochemical changes in the environment.
The LTFM allows the visualization and nonintrusive measurement of in situ concentrations of QDs in variably saturated porous media in real time across the whole experimental field. The LTFM allowed us to visualize and monitor QDs in variably saturated porous media and to acquire water saturations and in situ QD concentration measurements that are key parameters to provide insights into the mechanisms that govern the transport of nanoparticles in the subsurface environment. The ability to visualize QDs in porous media is significant, as the LTFM can provide both quantitative and qualitative data from the in situ research of nanoparticles in environmental matrices under various soil solution chemistry and hydrodynamic conditions.
The authors declare that there is no conflict of interests regarding the publication of this paper.
This work was supported primarily by the U.S. Department of Energy, Office of Science, Office of Biological and Environmental Research, Environmental Remediation Sciences Program under Grant no. DE-FG02-06ER64193. The authors thank Solidea Bonina for her technical assistance to this study.